Percutaneous transhepatic techniques for retrieving fractured and intrahepatically dislodged percutaneous transhepatic biliary drainage catheters
1 Department of Radiology, Taipei Veterans General Hospital Taipei, Taiwan  
Diagn Interv Radiol 2017; 23: 461-464
DOI: 10.5152/dir.2017.17064
Abstract

Dislodged intrabiliary drainage devices, including catheters, endoprostheses, and stents, may further impair drainage and cause various local reactions, vascular and gastrointestinal tract complications. Endoscopic approaches for management of plastic biliary endoprostheses have been extensively discussed. However, in rare cases of fracture of percutaneous transhepatic biliary drainage (PTBD) catheters, only a percutaneous transhepatic technique for retrieving should be applied to avoid further damage by its rigid fragment. We present the adjusted techniques using either a goose neck snare, over-the-wire balloon catheter, or biopsy forceps with image demonstration and reviews. We encountered two patients with PTBD tube fracture and intrahepatic dislodgment. In both patients, percutaneous approaches were used for successfully retrieving and removing the fractured catheter through transhepatic tract: one with the use of a biopsy forceps, another with an inflatable balloon catheter.

 

 

You may cite this article as: Hsien-Tzu L, Hsiuo-Shan T, Nai-Chi C, Yi-Yang L, Yi-You C, Chien-An L. Percutaneous transhepatic techniques for retrieving fractured and intrahepatically dislodged percutaneous transhepatic biliary drainage catheters. Diagn Interv Radiol 2017; 23:461–464.

Key Words
Authors
All
Author's Corner
Reviewer's Corner
Survey
Copyright © Turkish Society of Radiology | Latest Update: 01.11.2017