Diagnostic and Interventional Radiology
Interventional Radiology - Technical Note
Internal coil packing method for the Amplatzer vascular plug 4

Internal coil packing method for the Amplatzer vascular plug 4

1.

Department of Radiology, Kurume University School of Medicine, Kurume, Japan

Diagn Interv Radiol 2019; 25: 410-412
DOI: 10.5152/dir.2019.19009
Read: 274 Downloads: 40 Published: 07 October 2019

The Amplatzer vascular plugs (AVPs), including AVP 4, have excellent occlusion; however, insufficient embolization or recanalization may occasionally occur. This study aimed to evaluate the feasibility and effectiveness of internal coil packing method to insert microcoils into the AVP 4 using an experimental vascular model. The insertion of a 1.7 F microcatheter through the nitinol mesh of the AVP 4 was not possible. Only 0.010-inch electrolytic detachable microcoils could be inserted through catheter tips in contact with nitinol mesh. The internal coil packing of the AVP 4 may be accomplished by inserting 0.010-inch microcoils into the AVP 4 in cases of continued perfusion or recanalization with required short-segment embolization.,

 

You may cite this article as: Koganemaru M, Tanoue S, Kuhara A, Kugiyama T, Abe T. Internal coil packing method for the Amplatzer vascular plug 4. Diagn Interv Radiol 2019; 25:410–412.

Files
EISSN 1305-3612