Fast kilovoltage-switching dual-energy CT offering lower x-ray dose than single-energy CT for the chest: a quantitative and qualitative comparison study of the two methods of acquisition
1 Department of Radiology, Yeditepe University School of Medicine, İstanbul, Turkey  
Diagn Interv Radiol 2019; 25: 204-209
DOI: 10.5152/dir.2019.18412
Abstract

PURPOSE

We aimed to compare the size-specific dose estimates (SSDE), computed tomography (CT) dose indices and image quality parameters of the chest CTs obtained with fast kilovoltage-switching (FKS) dual-energy (DE) CT versus those with single-energy (SE) CT.

 

METHODS

Patients who had chest SECT within the last 6 months were prospectively scanned with chest FKS-DECT. Quantitative comparison was made by calculating the mean SSDE, CTDIvol, contrast, noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) for both acquisitions. Two radiologists evaluated the chest SECT and DECT images qualitatively blinded to the technique used. The paired Student’s t test was utilized for comparing the quantitative and qualitative data. Inter- and intraobserver agreement were also assessed.

 

RESULTS

A total of 42 patients were included. The mean SSDE, CTDIvol, contrast, noise, CNR, and SNR for SECT versus DECT were 12.7±2.2 mGy vs. 9.3±1.2 mGy (P = 0.001), 10.9±2.4 mGy vs. 8±1.2 mGy (P < 0.001), 211.9±44.7 vs. 216.3±59 (P = 0.350), 12.9±2.4 vs. 13.9±3.7 (P = 0.086), 13.5±5.2 vs. 13.3±8.4 (P = 0.548) and 12±3.5 vs. 11.5±3.4 (P = 0.774), respectively. Interobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.89, 0.85, 0.93, and 0.82, respectively; all P < 0.05). Intraobserver reproducibility was high for contrast, noise, CNR, and SNR (ICC = 0.80, 0.77, 0.85, and 0.88, respectively; all P < 0.05).

 

CONCLUSION

The mean SSDE of the chest CTs obtained with FKS-DECT were 26.8% lower than those with SECT with significant difference for the objective assessment and there was no significant difference for the subjective assessment of the image qualities, in this series.

 

 

You may cite this article as: Topçuoğlu OM, Sarıkaya B. Fast kilovoltage-switching dual-energy CT offering lower x-ray dose than single-energy CT for the chest: a quantitative and qualitative comparison study of the two methods of acquisition. Diagn Interv Radiol 2019; 25:204–209.